STM 3W Diode Laser Mode (continued)
Use the CW/P button to toggle between the CW (CONTINUOUS WAVE) AND PULSE modes.
Press the READY button in either mode to prepare the unit for Laser activation. A GREEN indicator will glow
next to the button identifying the “READY” status. Changing the mode or power level releases the unit from
“READY” status requiring the user to “SELECT” the new setting by pressing the READY button again. Setting
may not be changed when the laser is activated. In READY, the unit’s Aiming Beam is emitted.
The STM Diode Laser can be activated using either the handpiece or foot switch, but not both at the same time.
The footswitch, if attached, takes priority over the handpiece switch. Unplug the footswitch from the STM 3W
Diode Laser Unit if it will not be used.
When the ACTIVATOR button on the STM handpiece or the footswitch is depressed and held down, the active
mode indicator will turn from GREEN to RED and a audible “chirp” is produced, indicating that the Laser
working beam has been activated. The STM handpiece will then begin delivering laser emissions to the selected
tissue by means of the disposable 400 micron fiber delivery tip after a 200 micro-second delay. Release the
ACTIVATOR button or footswitch to deactivate the working power. Laser emissions will not be transmitted unless
a Fiber Optic Tip is fully inserted into the end of the Handpiece.
STAND-BY, CW Mode 0.8W
(default) output
STAND-BY, P Mode 0.8W
output
Laser in READY Mode @ 0.8W output