Gingivectomy, Gingivoplasty and Papillectomy with the Sapphire Portable Diode Laser:
Tip: Initiated
Mode: Continuous Wave
Power: 1.0 W
Anesthesia: Local, as required
Technique: Contact
1. Using a periodontal probe, record the depth of the pocket
to be treated and evaluate its condition, either acute or
chronically inflamed.
2. If there is no apparent exudate present, use the periodontal
probe to define the height of contour that is to be establish.
3. With the patient anesthetized, carefully place a stick mark along
the facial aspect of the gingiva to create a reference for the
incision path.
4. With the Fiber Optic Tip angled slightly toward the occlusal or
incisal, remove the diseased tissue down to the previously
marked incision line.
5. If necessary to achieve desired results, increase power level
but always use the lowest amount of power necessary to reach
the treatment objectives and avoid charring tissue.
6. Taper and festoon the crestal gingiva as needed and then
quickly remove the diseased epithelium and restore
anatomical contour.
Note: Remember to avoid touching the root whenever possible and
move quickly while near the bone or root.
7. Flush the pocket using a warm saline solution or
hydrogen peroxide.
8. Record power settings used and treatment times in the
patient’s chart.